The short version of stationary phase fits in a sentence. The long version — which is the one that helps — is below.
This page was last updated on 2025-01-23 and is reviewed periodically as new material appears.
Several separation modes exist, including reversed-phase, normal-phase, ion-exchange, size-exclusion, and hydrophilic interaction liquid chromatography. Reversed-phase uses a nonpolar stationary phase with a polar mobile phase and is widely applied to small organic molecules. Gradient elution changes mobile phase composition during the run, while isocratic elution keeps it constant. Column chemistry, particle size, temperature, flow rate, and mobile phase pH all influence retention and resolution. Method development selects conditions that separate analytes from matrix components and from each other.
Detection commonly uses ultraviolet-visible absorbance, fluorescence, refractive index, or mass spectrometry. Ultraviolet detection depends on molecular chromophores that absorb light at specific wavelengths. Mass spectrometry provides mass information and sensitive quantification, often after electrospray ionization. Before sample batches, performance checks examine resolution, elution time repeatability, peak symmetry, and plate count. Matrix effects and co-elution remain recognized uncertainties; formal validation studies and orthogonal detection help address them. Detector choice depends on analyte properties and required sensitivity.
Developing an HPLC test begins with defining the analytes, matrix, and required reporting limits. Chemists select a separation mode, column chemistry, mobile phase composition, flow rate, and detection wavelength or mass transition. Experiments then adjust these variables to achieve adequate retention, resolution, and peak shape. System suitability tests confirm that the instrument and method perform consistently before sample analysis. Without suitable resolution, quantitative results may be unreliable. Preliminary runs often use scouting gradients to locate retention windows.
Validation establishes that a method is suitable for its intended purpose. Typical parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, robustness, and stability of standards and samples. Acceptance criteria are defined in advance, and results are documented in a validation report. Regulatory guidance for pharmaceuticals, foods, and environmental testing differs, so the applicable framework must be identified. Ongoing verification uses control samples and trend charts after validation. Method transfer to another laboratory may require partial revalidation.
Routine quality control includes blanks, duplicates, spiked samples, and certified reference materials. Calibration curves are prepared with standards at several concentrations, and the detector response is checked for linearity. Carryover, column aging, mobile phase evaporation, and temperature drift can shift retention times or peak areas. Maintenance such as replacing seals, filters, and columns helps prevent failures. Records of injections, integration, and deviations support traceability. Audits may request raw data and instrument logs for each batch.
| Property | Value | Notes |
|---|---|---|
| Column particle size | 3–5 µm for conventional HPLC; sub-2 µm for UHPLC | Smaller particles increase backpressure and efficiency. |
| Typical flow rate | 0.5–2.0 mL/min for a 4.6 mm internal diameter column | Flow scales with column diameter and particle size. |
| UV detection wavelength | 190–400 nm | Selection depends on analyte chromophore. |
| Column temperature | 25–40 °C | Temperature affects retention, selectivity, and pressure. |
| Injection volume | 1–20 µL | Larger volumes may distort early-eluting peaks. |
HPLC testing separates dissolved compounds by passing a liquid sample through a column packed with stationary phase. A pump delivers mobile phase at controlled flow, and the sample components interact differently with stationary and mobile phases. Compounds that spend more time in mobile phase elute earlier; those retained by stationary phase elute later. Detectors record elution as peaks, and peak area or height relates to amount. This mechanism underpins quantitative analysis of mixtures.
Most routine HPLC testing uses reversed-phase columns, where the stationary phase is nonpolar and the mobile phase is a polar mixture such as water with an organic solvent. Analytes partition between the two phases according to polarity, size, and charge. Gradients that change solvent composition over time can separate compounds with broad retention ranges. Isocratic conditions keep solvent composition constant and suit simpler mixtures. The choice of column chemistry, pH, and temperature affects selectivity and peak shape.
Detection in HPLC testing commonly relies on ultraviolet-visible absorbance, fluorescence, refractive index, or mass spectrometry. UV detection is widely used because many organic compounds absorb light, but it requires a chromophore. Mass spectrometry provides mass-based identification and high sensitivity for trace analytes. Each detector has trade-offs in selectivity, cost, and compatibility with mobile phases. Quantification typically uses calibration curves prepared from reference standards. Results are reported as concentration, purity, or presence above a limit.
Method validation establishes that an HPLC procedure is suitable for its intended use. Key parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantitation, and robustness. Accuracy measures agreement with a true or accepted value, while precision describes repeatability and intermediate precision. Specificity confirms that the method measures the analyte without interference from impurities, degradants, or excipients. Validation is documented in a protocol and report, and acceptance criteria are set before experiments begin. Regulatory guidance varies by region, but the general principles are widely harmonized.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Common checks include retention time, peak area, resolution between critical pairs, tailing factor, and theoretical plate count. Results are compared with predefined limits, and a failed check requires investigation before sample results are reported. Quality control samples at low, middle, and high concentrations are injected at intervals to monitor accuracy and precision. Blank injections detect carryover and contamination, while control charts track performance over time.
Clinical trial number NCT02609776 for "Study of Amivantamab, a Human Bispecific EGFR and cMet Antibody, in Participants With Advanced Non-Small Cell Lung Cancer (CHRYSALIS)" at ClinicalTrials.gov Clinical trial number NCT04487080 for "A Study of Amivantamab and Lazertinib Combination Therapy Versus Osimertinib in Locally Advanced or Metastatic Non-Small Cell Lung Cancer (MARIPOSA)" at ClinicalTrials.gov Clinical trial number NCT04988295 for "A Study of Amivantamab and Lazertinib in Combination With Platinum-Based Chemotherapy Compared With Platinum-Based Chemotherapy in Patients With Epidermal Growth Factor Receptor (EGFR)-Mutated Locally Advanced or Metastatic Non- Small Cell Lung Cancer After Osimertinib Failure (MARIPOSA-2)" at ClinicalTrials.gov Clinical trial number NCT04538664 for "A Study of Combination Amivantamab and Carboplatin-Pemetrexed Therapy, Compared With Carboplatin-Pemetrexed, in Participants With Advanced or Metastatic Non-Small Cell Lung Cancer Characterized by Epidermal Growth Factor Receptor (EGFR) Exon 20 Insertions (PAPILLON)" at ClinicalTrials.gov
While genome annotation is primarily based on sequence similarity (and thus homology), other properties of sequences can be used to predict the function of genes. In fact, most gene function prediction methods focus on protein sequences as they are more informative and more feature-rich. For instance, the distribution of hydrophobic amino acids predicts transmembrane segments in proteins. However, protein function prediction can also use external information such as gene (or protein) expression data, protein structure, or protein–protein interactions. Evolutionary biology is the study of the origin and descent of species, as well as their change over time. Informatics has assisted evolutionary biologists by enabling researchers to:
Many of the payloads for oncology ADCs (oADC) are natural product based with some making covalent interactions with their target. Payloads include the microtubulin inhibitors monomethyl auristatin E (MMAE), monomethyl auristatin F (MMAF) and mertansine, DNA binder calicheamicin and topoisomerase 1 inhibitors SN-38 and exatecan resulting in a renaissance for natural product total synthesis. Glucocorticoid receptor modulators (GRMs) represent to most active payload class for iADCs. Approaches releasing marketed GRM molecules such as dexamethasone and budesonide have been developed. Modified GRM molecules have also been developed that enable the attachment of the linker with the term ADCidified describing the medicinal chemistry process of payload optimization to facilitate linker attachment. Alternatives to small molecule payloads have also been investigated, for example, siRNA. More recently, targeted protein degraders have been explored as payloads for antibody conjugates. A 2025 study described a KIF11-directed degrader–antibody conjugate (DAC), in which a cereblon-recruiting degrader was used as the payload to induce antigen-dependent protein degradation and cytotoxicity in preclinical models.
Sources: en.wikipedia.org
Quantitative studies of enzymatic reactions often found that the acceleration of chemical reaction speed cannot be fully explained by existing theories like the approximation, acid/base catalysis and electrophile/nucleophile catalysis. And there is an obvious paradox: in reversible enzymatic reaction if the active site perfectly fits the substrates then the backward reaction will be slowed since products cannot fit perfectly into the active site. So conformational distortion was introduced and argues that both active site and substrate can undergo conformational changes to fit with each other all the time.
Aberrant activation of Akt, either via PI3K or independently of PI3K, is often associated with malignancy. Studies have identified gene amplification of the Akt isoforms in many types of cancer, including glioblastoma, ovarian, pancreatic and breast cancers. Akt is also up-regulated in terms of mRNA production in breast and prostate cancer. Functional inactivation of PTEN, the major PI3K antagonist, can occur in cancer cells by point mutation, gene deletion or epigenetic mechanisms. Mutation in the pathway can also affect receptor tyrosine kinases, growth factors, Ras and the PI3K p110 subunit, leading to abnormal signaling activity. Therefore, many of the proteins in the pathway are targets for cancer therapeutics. In addition to its effects on cell survival and cell cycle progression, the PI3K-Akt pathway promotes other characteristics of cancer cells. Hyperactivity of the pathway promotes the epithelial-mesenchymal transition (EMT) and metastasis due to its effects on cell migration.
Developing standards-based messaging, to allow for automated data exchange Creating an exchange network, to ensure specific laboratories receive the information most pertinent to their work Improving reporting capabilities Selection and implementation of the best information management system for the laboratory's needs Strengthening the information network by sharing technical resources across public health laboratories One effort by the Informatics program is PHLIP, the Public Health Laboratory Interoperability Project. PHLIP aims to establish reliable laboratory data exchange between state public health laboratories and the CDC by fostering collaboration in IT and laboratory science.
Sources: en.wikipedia.org
Nycomed, a Swiss drug company which manufactures Actovegin, claims it can be used for circulation and nutrition disturbances, skin grafting, burns, and wound-healing impairment. Actovegin has also been used as a performance enhancer. It has been investigated for use in treatment of polyneuropathy in diabetes, and for stroke. One study found that when tissues suffer from hypoxia caused circulation abnormalities, Actovegin helps capillaries improve circulation by enhancing the neogenic mechanism in blood vessels. Anaphylactic shock has been observed in at least one patient treated with actovegin. There are reports suggesting that Actovegin have ergogenic ability, however one small trial found no apparent benefit from short-term use. Two studies certified Actovegin to have an anti-inflammatory effect. One of the two studies certified Actovegin to reduce Progressive muscular atrophy-induced inflammation in human cells. Another study certified Actovegin to have a positive impact on the growth of muscle cells.
Multivesicular bodies are endosomal compartments that sort ubiquitinated membrane proteins by incorporating them into vesicles. This process involves the sequential action of three multiprotein complexes, ESCRT I to III (ESCRT standing for 'endosomal sorting complexes required for transport'). Vps4p is a AAA-type ATPase involved in this MVB sorting pathway. It had originally been identified as a ”class E” vps (vacuolar protein sorting) mutant and was subsequently shown to catalyse the dissociation of ESCRT complexes. Vps4p is anchored via Vps46p to the endosomal membrane. Vps4p assembly is assisted by the conserved Vta1p protein, which regulates its oligomerization status and ATPase activity.
The kinetics of labeled derivatives of apamin were studied in vitro and in vivo in mice by Cheng-Raude et al. This shed some light on the kinetics of apamin itself. The key organ for excretion is likely to be the kidney, since enrichment of the labeled derivatives was found there. The peptide apamin is small enough to pass the glomerular barrier, facilitating renal excretion. The central nervous system, contrarily, was found to contain only very small amounts of apamin. This is unexpected, as this is the target organ for neurotoxicity caused by apamin. This low concentration thus appeared to be sufficient to cause the toxic effects. However, these results disagree with a study of Vincent et al. After injection of a supralethal dose of radioactive acetylated apamin in mice, enrichment was found in the spinal cord, which is part of the target organ. Some other organs, including kidney and brain, contained only small amounts of the apamin derivative. Symptoms following bee sting may include:
Sources: en.wikipedia.org
HPLC separates and quantifies compounds in a liquid sample. Detectors produce a response proportional to the amount of a compound passing through the flow cell. Identification by retention time requires comparison with a known standard.
UHPLC uses columns with smaller particles and operates at higher pressures than conventional HPLC. These conditions can improve speed, resolution, and sensitivity. Both techniques use the same fundamental separation principles.
Validation shows that a method performs reliably for its intended purpose across a defined range. It assesses accuracy, precision, specificity, linearity, and robustness. Regulated testing often requires documented validation before routine use.
System suitability is a set of checks performed before and during a run to confirm that the instrument, column, and method work as expected. Common checks include resolution, tailing factor, theoretical plates, and relative standard deviation of replicate injections. Failure triggers troubleshooting or method adjustment.